EasyManuals Logo

Honeywell DR4500A User Manual

Honeywell DR4500A
266 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #215 background imageLoading...
Page #215 background image
Output Calibration - Auxiliary Output Calibration
Release M DR4500A Classic Series Circular Chart Recorder With or Without Control Product Manual 201
April 2017
The procedure for calibrating the Auxiliary Output is listed in Table 8-5. Make sure “LOCKOUT” in the
Set Up group is set to “NONE.” See Section 3 Configuration. Also, “AUX OUT” must NOT be disabled.
Table 8-5 Auxiliary output calibration procedure
Step
Operation
Press
Action/Result
1
Enter Calibration Mode
SET
UP
Until you see:
Upper Display: CALIB
Lower Display: AUX OUT
2
Calibrate 0 %
FUNC
You will see:
Upper Display: (a value)
(approximately 750)
Lower Display: ZERO VAL
or
Until the desired 0 % (4 mA) output is read on the
milliammeter.
3
Calibrate 100 %
FUNC
Stores the 0 % value and you will see:
Upper Display: (a value)
(approximately 3650)
Lower Display: SPAN VAL
or
Until the desired 100 % (20 mA) output is read on the
milliammeter.
4
Exit the Calibration Mode
FUNC
The recorder stores the span value.
LOWR
DISP
or
SET
UP
To exit the calibration mode.
ATTENTION
The calibration for Auxiliary Outputs #2 and #3 follows the procedure for calibrating the current
proportional output (refer to Subsection 8.2).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell DR4500A and is the answer not in the manual?

Honeywell DR4500A Specifications

General IconGeneral
BrandHoneywell
ModelDR4500A
CategoryRecording Equipment
LanguageEnglish

Related product manuals