EasyManuals Logo

Honeywell DR4500A User Manual

Honeywell DR4500A
266 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #126 background imageLoading...
Page #126 background image
Configuration Parameter Definitions
Options Set Up Group
112 DR4500A Classic Series Circular Chart Recorder With or Without Control Product Manual Release M
April 2017
4.13 Options Set Up Group
Introduction
The Options Set Up Group lets you enable or disable some of the options that are available on the DR4500.
You can select the power frequency, select relative humidity as a measurement and a atmospheric pressure
compensation, select deviation recording action, and set the desired deviation setpoint.
Options group prompts
Table 4-13 lists all the function prompts in the Option set up group and their functions.
Table 4-13 Option group definitions
Lower Display
Prompt
Upper Display
Range of Setting or Selection
Parameter Definition
REJ FREQ
60HZ
50HZ
REJECTION FREQUENCY be sure that this selection
agrees with the line voltage frequency.
HF REJ
ENABLE
DISABL
HIGH FREQUENCY REJECTION This selection
allows high frequency noise rejection. It is used to quiet
noisy pens or PV.
ATTENTION
For fast processes it will cause
additional lag.
RELHUMID
NO
YES
RELATIVE HUMIDITY allows you to activate or
deactivate the RH calculation algorithm.
You must select 100RH input type for two inputs to
measure RH. It requires two 100-ohm RTD inputs so
INPUT 2 must be enabled for this prompt to appear.
Also, you must select RH for “PEN2IN” which
represents the “wet bulb” temperature when input 2 is
selected.
ATMPRES
590 to 800 mmHg
ATMOSPHERIC PRESSURE COMPENSATION This
prompt only appears if you select “YES” at “RELHUMID”
prompt previously. Select the desired compensation.
DEVIATION
NONE
SETPNT
CHAN1
DEVIATION selects deviation recording action.
NONE No deviation recording
SETPNT Records deviation from setpoint
CHAN1 Pen 1 records the difference between Input
1 and Input 2.
DEVSETPT
999.0 to 9999
DEVIATION SETPOINT VALUE sets the desired
deviation setpoint value for deviation recording.
This prompt only appears if “SETPNT” was selected
at prompt “DEVIATION” previously.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell DR4500A and is the answer not in the manual?

Honeywell DR4500A Specifications

General IconGeneral
BrandHoneywell
ModelDR4500A
CategoryRecording Equipment
LanguageEnglish

Related product manuals