EasyManua.ls Logo

Honeywell DR4500A - 5.8 Single Setpoint Ramp

Honeywell DR4500A
266 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation - Single Setpoint Ramp
Release M DR4500A Classic Series Circular Chart Recorder With or Without Control Product Manual 133
April 2017
5.8 Single Setpoint Ramp
Configuring the setpoint ramp
You can configure a single setpoint ramp to occur between the current local setpoint and a final local
setpoint over a time interval of from 1 to 255 minutes. You can RUN or HOLD the ramp at any time.
Procedure
75HTable 5-17 lists the procedure for configuring the Setpoint Ramp 1 or 2 parameters. The procedure for
Setpoint Program is in Section 6 Setpoint Ramp/Soak Programming Option.
Table 5-17 Procedure for configuring a single setpoint ramp
Step
Operation
Press
Action/Result
1
Select SP RAMP 1 or
SP RAMP 2 Set Up
group
SET
UP
Until you see:
Upper Display: SET UP
Lower Display: SP RAMP X
X = 1 or 2
2
Select the Setpoint Ramp
function
FUNC
Until you see:
Upper Display: DISABL
Lower Display: SP RAMP X
3
Enable Setpoint Ramp
Until you see:
Upper Display: ENABLE
Lower Display: SP RAMP X
NOTE: You cannot change the current local setpoint if
the setpoint ramp function is enabled. Make sure SP
PROG is disabled.
4
Set the Ramp Time
FUNC
Until you see:
Upper Display: The ramp time in minutes
Lower Display: TIME MIN
or
Change the upper display value to the number of
minutes in which you want the final setpoint to be
reached.
Setting Range = 1 to 255 minutes
NOTE: Entering “0” implies an immediate step change
to the final SP.
5
Set the Final Setpoint
value
FUNC
Upper Display: The final setpoint value
Lower Display: FINAL SP
6
or
Change the upper display value to the desired final
setpoint value.
Setting Range = within the setpoint limits

Table of Contents

Related product manuals