EasyManua.ls Logo

Honeywell DR4500A - 5.3 Start-up

Honeywell DR4500A
266 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation - Start-up
118 DR4500A Classic Series Circular Chart Recorder With or Without Control Product Manual Release M
April 2017
Set the Chart Time Line
54HTable 5-2 lists the steps required to set the chart time line.
Table 5-2 Set the chart time line procedure
Step
Press
Action/Result
1
Apply power and wait until the recorder runs its power-up tests.
2
CHART
To enable the chart drive function.
or
To drive the chart forward
To drive the chart backward
until current chart time line aligns with engraved time index on the chart
plate.
To increase forward drive speed, press and release the key while
holding in the key.
3
CHART
To disable the chart drive function and return to normal operation.
5.3 Start-up
Preparation
Before start-up, there are two tasks to perform:
Install or replace the circular chart as described in 55H5.16 56HRoutine Maintenance.
Be sure that the configured settings for the current time and wake time (time at which recording begins)
are correct. Note that the recorder is “awake” during start-up to verify interaction with the process.
During start-up, you should verify the interaction between your process and the DR4500A recorder to be
sure that your process can tolerate some upsets during this procedure.
Diagnostic tests
When power is applied, the recorder will run four diagnostic tests. 57HTable 5-3 lists the four diagnostic tests.
Table 5-3 Power up diagnostic tests
Prompt on Lower Display
Condition
RAM TEST
Check RAM
CONFTEST (configuration checksum)
Check nonvolatile memory
CAL TEST (working calibration)
Check calibration
BATTERY
Check the battery

Table of Contents

Related product manuals