ASAP 2460 Operator’s Manual Analysis Conditions Files
246-42800-01 - Aug 2013 3-25
pº and T button
Click to select options for obtaining the saturation pressure (p
o
) and
analysis bath temperature. Each selected option presents a different set of
parameters at the bottom of the screen.
Options and descriptions:
1. Analysis bath temperature (not used for krypton analyses).
Measures the p
o
on a continuous basis. Allows the measurement of
each data point without slowing the analysis.
2. Analysis bath temperature (not used for krypton analyses).
Measures the p
o
on a continuous basis and over the sample, then
adjusts the measured p
o
in the sample tube to agree with the p
o
over the
sample.
3. The p
o
and analysis bath temperatures or an analysis bath temperature
only (when Absolute pressure dosing is selected on the Analysis
conditions window). This method uses the entered values.
4. Psat gas
If this is a Krypton analysis, select the Krypton @ 77.35 K SOLID
option from the Psat Gas dropdown list.
Field or Button Description (continued)