Analysis Conditions Files ASAP 2460 Operator’s Manual
3-26 246-42800-01 - Aug 2013
pº and T button
(continued)
Click
Psat vs T to edit the values of the Psat vs T table. Editing the val-
ues in the current table does not affect the original table.
5. Analysis bath temperature. Analysis bath temperature and an estimate
for the initial p
o
(not used for krypton analyses).
Measures the p
o
over the sample. After the p
o
is measured, the value is
reapplied to all data points.
6. Analysis bath temperature.
This method calculates the p
o
at the time of analysis using the entered
temperature.
NOTE: If From last sample tube (for krypton analysis) on the analysis
window and a p
o
option other than 3, 4, or 6 is selected, an error message
is displayed when the analysis begins.
Field or Button Description (continued)