ASAP 2460 Operator’s Manual Report Options Files
246-42800-01 - Aug 2013 3-55
Horvath-Kawazoe Report Options
The Horvath-Kawazoe method plots individual peaks for different pore sizes even if the difference
between one pore size and the next is only one angstrom (0.10 nm) or less. In the Selected Reports list
box, highlight Horvath-Kawazoe, then click Edit.
Field or Button Description
Pore Geometry

group box
Select the option that best represents the physical geometry of the
micropores in the sample material. When Sphere is selected, options in the
Interaction Parameter group box are disabled.
Interaction Parameter
group box
Use to determine which interaction parameter will be used in the report.
These options are disabled if Sphere is selected in the Pore Geometry
group box.
• Computed - use to calculate using the parameters on the Horvath-
Kawazoe Physical Properties window (click the Properties button to
display the Physical Properties window). The interaction parameter is
recalculated each time a parameter in the Physical Properties window
is edited.
• Entered - select to calculate using the value entered in the text box.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Micromeritics ASAP 2460 and is the answer not in the manual?

Micromeritics ASAP 2460 Specifications

General IconGeneral
BrandMicromeritics
ModelASAP 2460
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals