ASAP 2460 Operator’s Manual Report Options Files
246-42800-01 - Aug 2013 3-57
Properties button
(continued)
Adsorptive group box:
Contains the parameters for the adsorptives (provided with the software
and/or user-defined). If using Computed for the interaction parameter, all
fields are enabled. If using Entered, only the values in the Diameter and
Diameter at zero energy text fields may be edited.
Mnemonic - select the mnemonic of the adsorptive gas in use.
Diameter - enter the diameter of the gas phase atom.
Diameter at zero energy - enter the diameter of an atom at zero
interaction energy: (2/5)
1/6
× diameter.
Polarizability - enter the polarizability of the adsorptive.
Magnetic susceptibility - enter the magnetic susceptibility of the
adsorptive.
Density - enter the density per unit area of the adsorptive.
Apply Cheng-Yang
correction checkbox
Use to apply the Cheng/Yang correction to the pore size analysis. This
correction substitutes the Langmuir equation of state for Henry’s Law in
the Horvath-Kawazoe derivation.
Smooth differentials
checkbox
Refer to BJH Adsorption/Desorption Report Options, page 3-49.
Selected Reports
list box
Select the types of reports to generate.
Cancel button
Edit button
OK button
Refer to Common Fields and Buttons - File Menu Options, page 3-1.