ASAP 2460 Operator’s Manual Start Analysis
246-42800-01 - Aug 2013 4-5
Field or Button Description
View dropdown list Refer to Common Fields and Buttons - Unit Menu Options,
page 4-2
Browse button Refer to Common Fields and Buttons - Unit Menu Options,
page 4-2
Port button Allows you to select the port to be used for the analysis. Only one
port may be selected at a time.
Sample text field Displays the sample’s file name.
Mass group box
ï€
(continued)
• Calculate - enables the Empty tube and Sample + tube
fields. Enter the values necessary to calculate the sample mass.
Equation used to calculate sample mass:
Mass
sample
= Mass
sample+tube
– Mass
tube
• Density - value is used for the Calculated free space method
only. Use 0.000 for a blank analysis.
p
o
text field Enabled if you have chosen to enter the p
o
measurement. Enter (or
confirm) the saturation pressure of the adsorptive, or edit the
current one if desired.