ASAP 2460 Operator’s Manual Appendix B
246-42800-01 - Dec 2013 B-7
2482- File cannot be opened for writing. It is already in use.
Cause: You are attempting to open a file that is currently being used (either by
this application or another).
Action: Locate the application using the file (in the Micromeritics applic
ation,
use the Windows menu item to get a list of all windows, one of which
may contain this file).
2483- An analysis cannot be performed on (file name). It is open for editing and con-
tains errors.
Cause: You attempted to use a sample file containing errors that is currently
open.
Action: Go to the window containing the file, correct the errors,
and save it.
2484- The edit session for (file name) must be saved before the analysis. Save
changes and continue with the analysis?
Cause: You attempted an analysis using a sample file that contains unsaved
changes.
Action: Select
Yes to save the changes and continue with the analysis.
Select
No to abort the analysis and return to the sample file.
2485- The service test file has an invalid status and cannot be used for this analysis.
Cause: The selected service test file has a status other than No Analysis.
Action: Select a different service test file or create a new one and use Replace
All to
copy parameters from the file you originally selected.
Cause C: A software error occurred when the file was accessed.
Action C: Contact your service representative.