Appendix B ASAP 2460 Operator’s Manual
B-6 246-42800-01 - Dec 2013
2478- Error copying sequential data segment.
Cause: An internal processing and/or hardware error occurred while accessing
a portion of a sample file.
Action: Confirm that the media being accessed does not contain errors; for
example,
you may wish to use a utility such as ScanDisk. Contact your
service representative if you continue to receive this error message.
2479- Cannot open (filename) because it is a (program) file.
Cause: You attempted to open a sample file (or parameter file) from another
application.
Action: Choose a sample file from the proper
application.
2480- File (file name) cannot be analyzed. It is currently being edited.
Cause: You attempted an analysis using a sample file that is being edited.
Action: Save the changes and close the dialog box.
2481- Error accessing the sample information file (file name (code #).
Action A: Run the Setup program and ensure that a valid port is specified; if not,
specify a valid one when prompted.
Action B: Reinstall the software, then re
start application.
Action C: Contact your service representative if
you continue to get this message.
Cause A: You attempted to open a file that is already open, possibly minimized.
Action A: View the minimized icons, locate and maximize the file.
Cause B: A computer or network problem occurred.
Action B: Check the performance of your computer devices or network.