ASAP 2460 Operator’s Manual Appendix B
246-42800-01 - Dec 2013 B-5
2471- Unit n - S/N; nn has an invalid communications port specified. It cannot be
initialized.
Cause: The communications port specified for this analyzer during installation
(or configuration) is invalid.
Action: Use the setup program to c
hange the analyzer configuration.
2474- (Unit {number}) Communications port COM (port number) specified in the pro-
gram control files is already in use. Unit cannot initialize.
Cause: The communications port assigned to the indicated unit is in use by
another program.
Action: End the program using the busy port or use the Setup program to
chang
e the communications port assigned to this program.
2475- (Unit {number}) Communications port COM (port number) specified in the pro-
gram control files cannot be accessed. Unit cannot initialize.
Cause: The operating system has prevented the application from accessing the
communications port.
Action: Review the hardware configuration of
the computer, ensure that no
other application is using the port. Contact your service representative
if you continue to receive this error message.
2476- Unit startup failed. Cannot initialize the communications port.
Cause: The communications port specified during installation is invalid.
Action: Use the setup program to change the ana
lyzer configuration. Contact
your service representative if you continue to receive this error mes-
sage.
2477- (Unit n; Serial nn) did not properly initialize.
Cause: The software was unable to initialize the analyzer.