Appendix B ASAP 2460 Operator’s Manual
B-4 246-42800-01 - Dec 2013
2459- An instrument is busy. A delay in restarting this application could result in loss
of new data. Continue with program Exit?
Cause: You attempted to exit the application while the analyzer was perform-
ing an operation.
Action: Rather than exiting the application you may choose to minimize it.
2460- Fatal Communications error on (unit nn).
Cause: Repeated attempts to maintain communication with the analyzer have
failed.
Action: Check the communications cable conn
ecting the analyzer to the com-
puter. Confirm that the analyzer is turned on and that the
status light is
not blinking.
If these checks are okay, exit the analysis program and then restart the
application. If
this error persists, contact your service representative.
2461- No instruments are in operation. This application will unconditionally
terminate.
Cause: At least one analyzer must be active for the application to operate. The
initialization of the analyzers configured with the Setup program has
failed. The application stops.
Action A: Usually this message is preceded by another me
ssage giving the reason
for the analyzer’s failure to initialize. See the instructions for that mes-
sage.
Action B: Check the cable connection between the analyzer and the computer.
V
erify that the analyzer has the power switch in the ON position and
that the light on the front panel is illuminated. If the application con-
tinues to fail in its attempts to initiali
ze the analyzer, contact your ser-
vice representative.
Action: Wait a few minutes before attempting to
exit the application again.