ASAP 2460 Operator’s Manual Appendix B
246-42800-01 - Dec 2013 B-3
2444- Not a valid file format.
2445- Subset wrote the wrong amount of data.
2446- Error reading data.
2447- Error writing data.
Cause: An unexpected error occurred when you tried to access a data file.
Action: Contact your service representative.
2448- File directory (path name) is invalid. Resetting to the installation directory.
Cause: A working directory specified in the .INI file is invalid. The directory
may have been deleted or moved to a different location.
Action: The installation directory will be substit
uted. The next time you open a
file, use the Directories: list to move to the correct directory.
2449- This field does not contain a valid file specification.
Cause: You entered an invalid file name.
Action: See the description of file naming conventions in your Windows man-
ual and re-enter the name.
2450- Sample Defaults may not be edited while this operation is in progress. Do you
wish to save and close the Sample Defaults edit session?
Cause: You are in the process of initiating an automatic analysis (an analysis in
which sample files are created using the defaults) while editing the
defaults.
Action: Finish your edit session of the defaults and close the dialog. Then
restart the automatic
analysis.
2458- An instrument is performing a critical operation. Wait a few moments before
exiting the application.
Cause: You attempted to exit the application while the analyzer is performing
a critical operation. This operation must be completed before the appli-
cation can be stopped.