Appendix B ASAP 2460 Operator’s Manual
B-28 246-42800-01 - Dec 2013
4057- At least two data points are needed for Freundlich calculations.
Cause: Less than two data points have been selected for the Freundlich report;
at least two are required.
Action: Select Freundlich points on the Coll
ected Data dialog. If calculation
assignments are not being used, edit the Freundlich Report options,
Absolute pressure range in the sample file.
4058- At least two data points are needed for Temkin calculations.
Cause: Less than two data points have been selected for the Temkin report; at
least two are required.
Action: Select Temkin points on the Collect
ed Data dialog. If calculation
assignments are not being used, edit the Temkin Report options, Abso-
lute pressure range in the sample
file.
Cause: You have selected at least one point with a negative pre
ssure value to
include in the Temkin report.
Action: Negative pressure points void Temkin calc
ulations. Open the sample
file and click the Collected Data tab. Locate the negative value(s) in
the Temkin column and deselect it (them).
4059- Fewer than 2 points available for MP-Method calculations.
Cause: At least two points are required for the MP-Method calculations.
Action: Edit the calculation assignments for the MP-Method report.
4060- Sample (file name) contains no data points.
Action: An attempt was made to save a sample without collected data as a t-
curve or alpha-S curve.
Cause: Repeat the sa
ve as t-curve or save as alpha-S operation after opening
a sample that has collected data.