4061- The t-curve must contain at least 2 points.
Cause: At least two points are required in a thickness curve definition.
Action: Edit the thickness curve.
4062- Error during report preparation.
Cause: An internal processing error has occurred.
Action A: Contact your Micromeritics service representative.
Action B: In Report Options, edit the report a
nd choose the desired subreport.
4063- No subreports selected.
Cause: A report was selected, but none of the subreports for that report were
selected, meaning nothing will be done for that report.
Action A: If the report was selected in error, remove the selection.
4067- No data points are within the range of pressures in the reference isotherm.
Cause: There are no collected data points within the range of pressures in the
reference isotherm.
Action: Select data points that
are in the range of the reference isotherm, or
select a more appropriate reference isotherm.
4068- No points were selected for the f-Ratio report.
Cause: The collected data column for the f-Ratio report does not have any
points selected.
Action: Edit the collected data dialog and select points for
the f-Ratio report.