EasyManuals Logo

Micromeritics ASAP 2460 User Manual

Default Icon
468 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #309 background imageLoading...
Page #309 background image
ASAP 2460 Operator’s Manual Appendix B
246-42800-01 - Dec 2013 B-29
4061- The t-curve must contain at least 2 points.
Cause: At least two points are required in a thickness curve definition.
Action: Edit the thickness curve.
4062- Error during report preparation.
Cause: An internal processing error has occurred.
Action A: Contact your Micromeritics service representative.
Action B: In Report Options, edit the report a
nd choose the desired subreport.
4063- No subreports selected.
Cause: A report was selected, but none of the subreports for that report were
selected, meaning nothing will be done for that report.
Action A: If the report was selected in error, remove the selection.
4067- No data points are within the range of pressures in the reference isotherm.
Cause: There are no collected data points within the range of pressures in the
reference isotherm.
Action: Select data points that
are in the range of the reference isotherm, or
select a more appropriate reference isotherm.
4068- No points were selected for the f-Ratio report.
Cause: The collected data column for the f-Ratio report does not have any
points selected.
Action: Edit the collected data dialog and select points for
the f-Ratio report.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Micromeritics ASAP 2460 and is the answer not in the manual?

Micromeritics ASAP 2460 Specifications

General IconGeneral
BrandMicromeritics
ModelASAP 2460
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals