Appendix B ASAP 2460 Operator’s Manual
B-30 246-42800-01 - Dec 2013
4069- Dosing method choice is invalid. The Analysis Conditions choice of Absolute
pressure dosing requires that the Adsorptive Properties (Dosing Method) is set to
Normal.
Cause: There is an incompatibility between the analysis conditions choice of
Absolute pressure dosing and the adsorptive properties dosing
method.
Action: Edit one of the choices.
4070- Unable to load deconvolution model (name).
Cause: For some reason, the list of available models was corrupted, therefore,
the model selected could not be loaded for the deconvolution.
Action: Exit the program and reinstall the software, then try again.
4071- The selected pressures points do not form a valid set for deconvolution.
Cause: The data points selected for analysis do not contain enough informa-
tion to allow a DFT data reduction.
Action A: Edit data points in the table of the C
ollected Data dialog, or select
another sample file. At least two points with strictly increasing pres-
sures and volumes adsorbed a
re required for a DFT Plus data reduc-
tion.
Action B: Edit the pressure range on the DFT report options dialog.
4072- The range of pressures selected is too small to deconvolute using this model.
Cause: A null result was found using the selected model.
Action: Edit data points in the table of the C
ollected Data dialog, or select
another sample file. At least two points with strictly increasing pres-
sures and volumes adsorbed
are required for a DFT data reduction.