ASAP 2460 Operator’s Manual Appendix B
246-42800-01 - Dec 2013 B-31
4073- The analysis gas (name) does not match the model gas (name).
Cause: The model assumes a specific gas and the sample file uses a different
one.
Action: Select a model that assumes the same gas.
4074- The analysis temperature (value) does not match the model temperature
(value).
Cause: The wrong model was selected for DFT reporting.
Action: Select a model for which the analysis temperature matches.
4075- The models cannot be located in the models folder. Reinstall the software.
Cause: There are no models in the MODELS subdirectory (of the program
directory). These are placed in the directory during software installa-
tion.
Action: Reinstall the software to place the DFT models in the MOELS direc-
tory.
4076- Invalid sample file - Adsorptive Properties “Non-condensing adsorptive” can-
not be used unless Analysis Conditions “Absolute pressure dosing” is selected.
Cause: You selected a sample file that has Non-condensing adsorptive
selected in the Adsorptive Properties but does not have Absolute pres-
sure dosing selected in Anal
ysis Conditions.
Action A: If you plan to use a non-condensing adsorptive, select Absolute
pres-
sure dosing on
the Analysis Conditions dialog.
Action B: If you are not using a non-condensing ads
orptive, deselect the Non-
condensing adsorptive option on the Adsorptive Properties dialog, or
choose an adsorptive that you are sure in not non-condensing.