ASAP 2460 Operator’s Manual Appendix B
246-42800-01 - Dec 2013 B-43
6522- Time limit exceeded while evacuating manifold.
Cause A: Maximum manifold evacuation time was exceeded before the vacuum
set point was achieved. Vacuum pump may be turned off.
Action A: Turn on vacuum pump switch. Then restart the analysis.
Cause B: The vacuum pump oil level is low.
Action B: Check the vacuum pump oil level and a
dd more oil if necessary. Then
restart the analysis.
Cause C: The manifold is contaminated or leaking.
Action C: Correct the problem. Refer to Appendix D, TESTING THE ANA-
LYZER FOR LEAKS.
6523- Analysis suspended: Time limit exceeded while evacuating sample
(unrestricted).
Cause: The maximum time for evacuating the sample through the unrestricted
valve was exceeded.
Action: Check the sample tube and the sample tube fit
ting; ensure that the tube
is securely attached to the port. Verify that the sample is properly
degassed; then restart the analysis.
Action A: Set the analysis gas regulator to 15
psig (1.0 bar). Then resume the
analysis.
Cause B: The analysis gas bottle is empty.
Action B: Connect a new analysis gas bottle. Then
resume the analysis.