EasyManuals Logo

Micromeritics ASAP 2460 User Manual

Default Icon
468 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #324 background imageLoading...
Page #324 background image
Appendix B ASAP 2460 Operator’s Manual
B-44 246-42800-01 - Dec 2013
6524- Analysis suspended: Time limit exceeded while evacuating sample (servo).
Cause: The maximum time for evacuating the sample through the restricted
valve was exceeded. Possible causes are a leak in the sample tube fit-
ting, a crack in the sample tube,
or a poorly degassed sample.
Action: Check the sample tube and the sample tube fitting; ens
ure that the tube
is securely attached to the port. Verify that the sample is properly
degassed; then resume the analysis.
6525- Time limit exceeded while backfilling manifold to (target pressure) with (gas).
Cause A: The maximum time was exceeded before the target pressure point was
reached. The gas regulator may be set too low or turned off.
Action A: Set the gas regulator to 15 psig (1.03 bar). Then res
ume the analysis.
Cause B: The gas bottle is empty.
Action B: Connect a new gas bottle. Then resume
the analysis.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Micromeritics ASAP 2460 and is the answer not in the manual?

Micromeritics ASAP 2460 Specifications

General IconGeneral
BrandMicromeritics
ModelASAP 2460
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals