6 Maintenance and Troubleshooting
Field or Button Description
Servo Valve. Open.
Sample Tube. Cannot be manually controlled.
Smart VacPrep Schematic Components Icon Table (continued)
REPLACE THE SMART VACPREP FUSE
The Smart VacPrep uses two 5 amp fuses located in a compartment adjacent to the power connector on
the back of the unit.
1. Use a flat head screwdriver to gently pry open the cover of the compartment.
2. Use the flat head screwdriver to gently pry the fuses out of the holder.
3. Insert a new fuse into each holder, then push the holders back into the slot until snug.
4. Press the compartment cover shut. Ensure it closes securely.
SAMPLE TUBE INSTALLATION ON THE SMART VACPREP
1. While holding the degas port plug, turn the connector nut counterclockwise to loosen. Remove
both the plug and the nut.
2. If using a rubber stopper, remove it from the sample tube.If using a Check Seal or TransSeal, do
not remove them from the sample tube. Place the port connector nut, ferrule, and O-ring onto the
sample tube.
6 - 2 Smart VacPrep Operator Manual