11 - 8
1. Prepare the patient and place the cuff as instructed in 11.6 Preparing for NIBP
Measurement.
2. Select a workflow under Continuous Monitoring.
3. Select to set the patient position and measurement site.
4. Select the NIBP numeric area to enter the NIBP menu.
5. Set Interval to an appropriate duration.
6. Set Start Mode to Interval or Clock.
Clock: after the first measurement, the monitor automatically synchronizes
NIBP automatic measurement with the real time clock. For example, if
Interval is set to 20 min, and you start NIBP auto measurement at 14: 03, the
next measurement will be taken at 14: 20, and then at 14:40, 15:00, and so
on.
Interval: after the first measurement, the monitor automatically repeats
measurement at set interval. For example, if Interval is set to 20 min, and
you start NIBP auto measurement at 14:03, the next measurement will be
taken at 14:23, and then at 14:43, 15:03, and so on.
7. Select Start NIBP.
11.7.3 Performing STAT Measurement
To perform STAT measurement, follow this procedure:
1. Prepare the patient and place the cuff as instructed in 11.6 Preparing for NIBP
Measurement.
2. Select a workflow under Continuous Monitoring.
3. Select to set the patient position and measurement site.
4. Select the NIBP numeric area to enter the NIBP menu.
5. Select STAT.
11.7.4 Performing Sequence Measurement
To perform sequence measurement, follow this procedure:
1. Prepare the patient and place the cuff as instructed in 11.6 Preparing for NIBP
Measurement.
2. Select a workflow under Continuous Monitoring.
3. Select to set the patient position and measurement site.
4. Select the NIBP numeric area to enter the NIBP menu.
5. Select Sequence tab to set the duration and interval of phases in the sequence.For
details, see 11.8.4 Setting NIBP Sequence.
6. Select Setup tab, set Interval to Sequence.