B EMC and Radio Regulatory Compliance
B.1 EMC
The device meets the requirements of IEC 60601-1-2: 2014.
• Use of accessories, transducers and cables other than those specified or 
provided by the manufacturer of this device could result in increased 
electromagnetic emissions or decreased electromagnetic immunity of this 
device and result in improper operation.
• The non-ME EQUIPMENT (e.g. ITE) that is a part of an ME SYSTEM may be 
disrupted by the electromagnetic interference of nearby equipment. It may 
be necessary to take mitigation measures, such as re-orienting or relocating 
the non-ME EQUIPMENT or shielding the location.
• Use of this device adjacent to or stacked with other device should be avoided 
because it could result in improper operation. If such use is necessary, this 
device and the other device should be observed to verify that they are 
operating normally. 
• This device is intended for use in professional healthcare facility EMC 
environment. If it is used in special environment, such as magnetic resonance 
imaging environment, the equipment/system may be disrupted by the 
operation of nearby equipment.
• Portable RF communications equipment (including peripherals such as 
antenna cables and external antennas) should be used no closer than 30 cm 
(12 inches) to any part of the this device, including cables specified by the 
manufacturer. Otherwise, degradation of the performance of this device 
could result.