B - 1
B EMC and Radio Regulatory Compliance
B.1 EMC
The device meets the requirements of IEC 60601-1-2: 2014.
NOTE
Use of accessories, transducers and cables other than those specified or
provided by the manufacturer of this device could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this
device and result in improper operation.
The non-ME EQUIPMENT (e.g. ITE) that is a part of an ME SYSTEM may be
disrupted by the electromagnetic interference of nearby equipment. It may
be necessary to take mitigation measures, such as re-orienting or relocating
the non-ME EQUIPMENT or shielding the location.
Use of this device adjacent to or stacked with other device should be avoided
because it could result in improper operation. If such use is necessary, this
device and the other device should be observed to verify that they are
operating normally.
This device is intended for use in professional healthcare facility EMC
environment. If it is used in special environment, such as magnetic resonance
imaging environment, the equipment/system may be disrupted by the
operation of nearby equipment.
Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30 cm
(12 inches) to any part of the this device, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this device
could result.