20 - 5
20.4 Cleaning and Disinfecting the Accessories
For the NIBP air hose and SpO
2
cable, you should clean and disinfect them using the
cleaners and disinfectants and methods listed in this section. For other accessories, you
should consult the instructions delivered with the accessories.
• Fluids entering the NIBP air hose can damage the equipment. When cleaning
or disinfecting the NIBP air hose, prevent liquid from entering the hose.
• Periodically inspect the NIBP air hose and connector for signs of wear or
deterioration after cleaning or disinfecting the NIBP air hose. Replace the
NIBP air hose if you detect a leak. Dispose of damaged NIBP air hose
according to local laws for disposal of hospital waster.
• Never immerse or soak the accessories in any liquid.
• Never clean or disinfect the connectors and metal parts.
• Use only Mindray approved cleaners and disinfectants and methods listed in
this section to clean or disinfect the accessories. Warranty does not cover
damage caused by unapproved substances or methods.
• To avoid long term damage, the accessories should be disinfected only when
necessary as determined by your hospital’s policy.
Descosept® forte
Liquid Dr. Schumacher GmbH
Descosept® AF
Liquid Dr. Schumacher GmbH
Dismozon® plus, 0.4%
Powder BODE Chemie GmbH
mikrozid® AF Wipes
Wipes Schülke & Mayr GmbH
Terralin® Liquid
Liquid Schülke & Mayr GmbH
Perform® Classic
Concentrate OXY, 0.5%
Powder Schülke & Mayr GmbH
S-7 hospital grade disinfectant
surface wipes
Wipes ANAERON
Neutral detergent wipes Wipes RAYNARD Health
Premier disinfectant wips Wipes RAYNARD Health
V-wipes Wipes Whiteley Medical
Product Name Product Type Manufacturer