21 - 4
If your monitor needs a safety test and performance test, contact the service personnel.
21.5.1 Performing Visual Inspection
Visually inspect the equipment before its first used every day. If you find any signs of
damage, remove your monitor from use and contact the service personnel.
Verify that the equipment meets the following requirements:
Environment and power supply specifications are met.
The monitor housing and display screen are free from cracks or other damages
The power cord is not damaged and the insulation is in good condition.
Connectors, plugs, and cables are not damaged and kinked.
Power cord and patient cables are securely connected with the equipment.
21.5.2 Performing Power-on Test
The monitor automatically performs a selftest at startup. Check the following items for
the power-on test:
The equipment powers on properly.
The alarm system works properly.
The monitor displays properly.
21.5.3 Testing the Recorder
To test the recorder, follow this procedure:
1. Start a recording task to print waveforms and reports.
2. Check that the recorder functions correctly.
3. Check that the printout is clear without missing dots.
21.5.4 Checking the Battery
For information on battery check, see 19.8.2 Checking Battery Performance.
21.6 Disposing of the Monitor
Dispose of the monitor and its accessories when its service life is reached. Follow local
regulations regarding the disposal of such product.
WARNING
For disposal of parts and accessories, where not otherwise specified, follow
local regulations regarding disposal of hospital waste.