D - 3
D.2.1 General Technical Alarm Messages
Note:XX represents a measurement or parameter label, such as SpO
2
, NIBP, CO
2
, and so on.
D.2.2 SpO
2
Technical Alarm Messages
Alarm message
Default
priority
Indication on
alarm reset
Cause and solution
XX Module Error High C XX module does not work properly.
Contact your service personnel.
Alarm message
Default
priority
Indication on
alarm reset
Cause and solution
SpO2 Sensor Off Low B The SpO
2
sensor has become
detached from the patient or the
module. Check the sensor
connection. If the alarm persists,
replace the sensor.
SpO2 No Sensor Low A The SpO
2
extension cable is
detached from the SpO
2
module, or
the SpO
2
sensor is detached from
the SpO
2
extension cable. Check the
SpO
2
cable and the sensor
connection. If the alarm persists,
replace the sensor.
SpO2 Excess
Light
Low C Ambient light is too strong. Move the
sensor to a place with lower level of
ambient light or cover the sensor to
minimize the ambient light.
SpO2 No Pulse Low C The SpO
2
sensor failed to obtain
pulse signal. Check the patient’s
condition and replace the sensor
application site. If the alarm persists,
replace the sensor.
SpO2 Sensor
Incompatible
Low C Incompatible or an unspecified SpO
2
sensor is used. Use specified sensors.