D - 4
SpO2 Low Signal 
Quality
Low C 1. Check the sensor and sensor 
position.
2. Make sure the patient is not 
shivering or moving.
3. The patient’s pulse may be too low 
to be measured.
SpO2 
Interference
Low C The SpO
2
 signal has been interfered. 
Check for any possible sources of 
signal noise and check the patient 
for excessive motion.
SpO2 Sensor 
Error
Low C Replace the sensor and measure 
again.
SpO2 Searching 
Pulse
Prompt / SpO
2
 is searching for pulse.
SpO2 Low 
Perfusion
Prompt / The SpO
2
 sensor is not properly 
placed or the patient’s perfusion 
index is too low.
1. Check the sensor and sensor 
position.
2. Reposition the sensor if necessary.
Replace SpO2 
Cable
Low C The SpO
2
 cable expired, cannot be 
recognized, or failed. Replace the 
SpO
2
 cable.
Prompt The SpO
2
 cable is about to expire.
Replace the SpO
2
 cable.
Incompatible 
SpO2 Cable
Low C The SpO
2
 cable connected is not 
compatible with the equipment. 
Replace the SpO
2
 cable.
Replace SpO2 
Sensor
Low C The SpO
2
 sensor expired. Replace 
the SpO
2
 sensor.
Prompt The SpO
2
 sensor is about to expire.
Replace the SpO
2
 sensor.
Check SpO2 
Sensor 
Connection
Low C No SpO
2
 sensor is connected. 
Connect an SpO
2
 sensor.
Alarm message
Default 
priority
Indication on 
alarm reset
Cause and solution