21 - 3
21.4 Checking Version Information
You may be asked for information on monitor and module version. 
To view system software version information, select the Main Menu quick key →from 
the System column select Version.
You can also view more version information by following this procedure
1. Select the Main Menu quick key → from the System column select Maintenance 
→ input the required password → select .
2. Select the Version tab.
You can check system software version, module hardware and software version, and 
firmware version.
21.5 Testing Methods and Procedures
Except the following maintenance tasks, all other test and maintenance tasks should be 
performed by Mindray-qualified service personnel only. 
■ Regular check, including visual inspection and power-on test
■ Recorder test
■ Battery check
Nurse call test If you suspect that the nurse call function does not work 
properly.
Electrical Safety Tests
Electrical safety tests 
Once every two years. 
Other Tests
Power-on test  Before use.
Recorder check 1. When the recorder is used for the first time.
2. Follow any repair or replacement of the recorder.
Battery 
check 
Functionality 
test 
1. When first installed. 
2. When battery is replaced. 
Performance 
test 
Every three months or if the battery runtime reduced 
significantly. 
Test/Maintenance Item  Recommended Frequency 
Performance Tests