18 - 7
NOTE
The alarm volume escalation function is not applied to the latched alarms.
When your monitor is connected to the CMS and Minimum Alarm Volume is
set to 0, Minimum Alarm Volume automatically changes to 2 if the CMS is
disconnected.
18.6.2 The Pause/Reset Tab
Auto Increase
Volume
2 Steps 2 Steps: if an alarm is not reset within the
designated delay time after the alarm
occurs, the alarm volume automatically
increases by two levels.
1 Step: if an alarm is not reset within the
designated delay time after the alarm
occurs, the alarm volume automatically
increases by one level.
Off: if an alarm is not reset within the
designated delay time after the alarm
occurs, the volume of the alarm tone does
not change.
Increase Volume
Delay
20 sec Defines the delay time of alarm volume
escalation
Defaults / Restores default settings on current page.
Menu Item Default Setting Function
Section Menu Item
Default
Setting
Function
Pause Pause Alarm
Pause
Selects the pause function.
Alarm Pause: pauses alarms.
Audio Pause: pauses alarm tones.
Pause Time 2 min Selects the alarm pause time.
The alarm pause time can be set to 1 min, 2
min, 3 min, or Permanent.