D - 9
D.2.7 Power Supply Technical Alarm Messages
Alarm message
Default
priority
Indication on
alarm reset
Cause and solution
Low Battery Med C Connect the monitor to an AC
power source and allow the
batteries to charge.
Critically Low
Battery
High C Connect the monitor to an AC
power source and allow the
batteries to charge.
Battery Service
Required
Low B The battery reaches its lifetime.
Replace the battery.
Power Board
Comm Error
High C Restart the monitor. If the alarm
persists, contact your service
personnel.
Battery Error High C The battery may fail. Contact your
service personnel.
Battery Charging
Error
High C The charging circuit fails or the
battery fails. Contact your service
personnel.
Battery
Temperature Too
High
High C Stop using the monitor after this
alarm appears, and contact your
service personnel.
Battery Off High C Restart the monitor. If the alarm
persists, contact your service
personnel.
RT Clock Need
Reset
High C Contact your service personnel.
RT Clock Not Exist High C Contact your service personnel.