7 - 7
2. Select the Limits tab.
3. On the Limits page, select Defaults at the bottom.
7.6.6 Setting the CMS and eGateway Disconnection Alarm
You can choose whether to issue an alarm when the monitor is not connected or
disconnected from the CMS/eGateway, and set the priority of these alarms.
This setting is password protected. For details, see 18.6.5 The Other Tab.
7.6.7 Setting the Switch of the SpO
2
Desat Alarm Off
You can choose whether switching off the SpO2 Desat alarm is permissible or not. This
setting is password protected. For more information, see 18.6.5 The Other Tab.
WARNING
If you switch off the SpO2 Desat alarm, the monitor will not alarm when the
patient’s SpO
2
is extremely low. This may result in a hazard to the patient.
Always keep the patient under close surveillance.
7.6.8 Setting the Switch of the Apnea Alarm Off
You can choose whether switching off the apnea alarm is permissible or not. This setting
is password protected. For more information, see 18.6.5 The Other Tab.
WARNING
If you switch off the apnea alarm, the monitor will not issue the apnea alarm
in case that apnea happens. This may result in a hazard to the patient. Keep
the patient under close surveillance.
7.7 Pausing Alarms/Pausing Alarm Tones
7.7.1 Defining the Pause Function
You can either pause alarms or pause alarm tones. This depends on the pause
setting.This setting is password protected. For more information, see 18.6.2 The Pause/
Reset Tab.