11 - 11
■ Patient having frequent arrhythmia
■ Weak patient pulse
• The monitor always performs NIBP measurement at deflation stage if Patient
Category is Neo.
11.8.8 Setting Measurement Times
To set the measurement times, follow this procedure:
1. Select the Main Menu quick key → from the Configuration column select
Workflow → input the required password → select .
2. Select on the right of the workflow to be set. Make sure the workflow is
under Spot Check.
3. Select Parameters Setup.
4. Under NIBP tab, set Measurements to Twice.
11.8.9 Correcting the NIBP Measurements
The middle of the cuff should be at the level of right atrium. If the limb is not at the heart
level, you need to correct the measurement:
■ Add 0.75 mmHg (0.10 kPa) to the displayed value for each centimetre higher.
■ Deduct 0.75 mmHg (0.10 kPa) to the displayed value for each centimeter lower.
11.9 BP Averaging
In BP averaging mode, a group of BP measurement is performed on a patient
automatically. At the completion of measurement, a group of averaged BP readings is
displayed.
11.9.1 Enabling BP Averaging
BP averaging is available only when the Department is set to Physician Office and the
monitor is working in Spot Check mode.
To change the department, see 5.2 Changing the Department.
To enable BP averaging, follow this procedure:
1. Select the Main Menu quick key → from the Configuration column select
Workflow → input the required password → select .
2. Select on the right of the workflow to be set. Make sure the workflow is
under Spot Check.
3. Select Parameters Setup → the NIBP tab.