D - 5
D.2.3 Temp Technical Alarm Messages
SpO2 No Cable
Connected
Low C No SpO
2
cable is connected. Connect
an SpO
2
cable.
Alarm message
Default
priority
Indication on
alarm reset
Cause and solution
Alarm message
Default
priority
Indication on
alarm reset
Cause and solution
Cannot Take
Neonatal Rectal
Temp (for
SmarTemp)
High B Rectal measurement is not
applicable for neonatal patient.
Select another site.
Temp No Probe
(for SmarTemp)
Med B The Temp probe is disconnected or
damaged. Reconnect the probe. If
the problem persists, change for
another probe.
Temp Probe
Misplaced (for
SmarTemp™)
Med A The measured value is invalid.
Replace the probe and check the
value.
Temp Measuring
Timeout (for
SmarTemp™)
Med B The measurement has exceeded
regular measurement time. Try
another measurement.
Ambient Temp
Overrange (for
SmarTemp™)
Low A The ambient temperature is too low.
Move the patient to a warmer place
and take temperature measurement
again if necessary.
Temp Overrange
(for SmarTemp™)
Low C The temperature measurement
exceeds the measurement range.
Check the patient’s condition.
Thermal Resistor
Err (for
SmarTemp™)
Med C The Temp module may fail. Take
temperature measurement again. If
the alarm persists, contact your
service personnel.
TempIF
Thermometer
Error (for
Genius™ 3)
High C The thermometer may fail. Take
temperature measurement again. If
the alarm persists, replace the
thermometer.