18 - 10
18.6.5 The Other Tab
Alarm Priority High Only Selects the priority of alarms sent to the nurse call
system
Alarm Type Physiological
Only
Selects the type of alarms sent to the nurse call
system.
Menu Item Default Setting Function
Section Menu Item
Default
Setting
Function
Alarm Priority SpO2 Sensor
Off
Low Selects the alarm level for SpO
2
sensor off
alarm.
CMS/eGW
Disconnected
Low Selects the priority of the CMS and
eGateway disconnection alarm.
This setting is available only in
Continuous Monitoring mode.
Alarm Delay Alarm Delay 6 sec 1 sec - 15 sec for continuously
measured parameters, the monitor
does not present the alarm if the alarm
condition is resolved within the delay
time.
Off: an alarm is always presented.
The setting of Alarm Delay is not applied
to the apnea alarms.
Other SPO2 Desat
Alarm Off
Disable Selects whether the SpO
2
Desat alarm
can be switched off.
Disable: the SpO
2
Desat alarm cannot
be switched off.
Enable: the SpO
2
Desat alarm can be
switched off.
Apnea Alarm
Off
Disable Selects whether the apnea alarm can be
switched off.
Disable: the apnea alarm cannot be
switched off.
Enable: the apnea alarm can be
switched off.