C - 5
C.1.5 CO
2
Default Settings
Item Default Setting
Alarm RR On/Off On
High Adult and pediatric: 30 rpm
Neonate: 100 rpm
Low Adult and pediatric: 8 rpm
Neonate: 30 rpm
Priority Med
Apnea On/Off On (Locked)
Priority High
EtCO
2
On/Off On
High Adult and pediatric: 50 mmHg
Neonate: 45 mmHg
Low Adult and pediatric: 25mmHg
Neonate: 30 mmHg
Priority Med
FiCO
2
On/Off On
High 4 mmHg
Priority Med
Apnea Delay Adult and pediatric: 20 sec
Neonate: 15 sec
Setup Operating Mode Measure
Auto Standby 60 min
BTPS Compensation Off
O2 Compensation 21%
N2O Compensation 0%