20 - 8
20.4.3 Cleaning the Accessories
You should clean the accessories (NIBP air hose, and SpO
2
cable) on a regular basis.
Before cleaning the accessories, consult your hospital’s regulations for cleaning the
accessories.
To clean the accessories (NIBP air hose and SpO
2
cable), follow this procedure:
1. Clean the accessories with a soft cloth moistened with water or ethanol (70%).
2. Wipe off all the cleaner residue with a dry cloth.
3. Allow the accessories to air dry.
20.4.4 Disinfecting the Accessories
We recommend that the accessories (NIBP air hose and SpO
2
cable) should be
disinfected only when necessary as determined by your hospital’s policy. Cleaning the
accessories before disinfecting is recommended.
20.5 Sterilization
Sterilization is not recommended for this monitor, related products, accessories, or
supplies unless otherwise indicated in the Instructions for Use that accompany the
products, accessories or supplies.
20.6 Impact of Improper Cleaning
Using cleaners other than those recommended may have the following impact:
Product discoloration
Metal part corrosion
Brittle and breaking wires, connectors, and equipment housing
Reduced cable and leadwire life
Overall system performance degradation
Equipment malfunction or failure