D - 8
D.2.5 CO
2
Technical Alarm Messages
D.2.6 EWS Technical Alarms
Note: XX represents RR, SpO2, Supp. O2, Temp, BP, Consciousness, Blood Sugar etc.
Alarm message
Default
priority
Indication on
alarm reset
Cause and solution
CO2 Module
High Temp
Low C Ambient temperature is too high or
there is a module failure.
1. Lower the operating temperature.
2. If the alarm persists, the CO
2
module may fail, contact your service
personnel.
CO2 Zero Failed Low C Try to zero the module again. If the
alarm persists, contact your service
personnel.
CO2 Airway
Occluded
Low C 1. Check if the sample line is kinked
or occluded.
2. Replace the sample line.
3. If the alarm persists, contact your
service personnel.
XX Overrange Low C The measured value exceeds the
module measurement range. Check
the patient’s condition.
CO2 No Adaptor Low B Check the adaptor connection.
Alarm message
Default
priority
Indication on
alarm reset
Cause and solution
EWS param XX is
timeout
Low A The manually input parameter is
timeout. Input a parameter numeric
again.
EWS score needs
to be confirmed
Low A Confirm to save or give up current
score.