A - 1
A Product Specifications
A.1 Classifications
The equipment is classified, according to IEC60601-1:
A.2 Environmental Specifications
A.2.1 Main Unit
Type of protection against
electrical shock
CLASS I EQUIPMENT, equipment energized from an
external and internal electrical power source.
Degree of protection against
electrical shock
DEFIBRILLATION-PROOF TYPE CF AAPPLIED PART for
SpO
2
, NIBP, and SmarTemp™ Temp
TYPE BF APPLIED PART for Genius™ 3 Temp
Type B APPLIED PART for HeTaida Thermometer
DEFIBRILLATION-PROOF TYPE BF APPLIED PART for CO
2
and Exergen TemporalScanner ™ Temp.
Mode of operation Continuous
Degree of protection against
harmful ingress of water
IPX2 (Protection against vertically falling water drops
when ENCLOSURE tilted up to 15°)
Degree of safety of application
in the presence of a
FLAMMABLE ANAESTHETIC
MIXTURE WITH AIR or WITH
OXYGEN OR NITROUS OXIDE
EQUIPMENT not suitable for use in the presence of a
FLAMMABLE ANAESTHETIC MIXTUREWITH AIR or WITH
OXYGEN OR NITROUS OXIDE
Degree of mobility Portable
Item Operating conditions Storage conditions
Temperature 0 to 40 ºC (without CO
2
)
5 to 40 ºC (with CO
2
)
–20 to 60 ºC
Relative humidity (non-
condensing)
15% to 95% 10% to 95%
Barometric 57.0 to 107.4 kPa (without CO
2
)
57.3 to 105.3 kPa (with CO
2
)
16.0 to 107.4 kPa
(without CO
2
)
57.3 to 105.3 kPa (with
CO
2
)