11 - 13
11.9.4 Changing BP Averaging Settings
11.9.4.1 Setting the Delay before Starting the First Measurement
Follow this procedure:
1. Select the Main Menu quick key → from the Configuration column select
Workflow → input the required password → select .
2. Select on the right of the workflow to be set. Make sure the workflow is
under Spot Check.
3. Select Parameters Setup → the NIBP tab.
4. Under Average, set Delay Before Starting Measurement.
11.9.4.2 Setting the Number of Measurement
Follow this procedure:
1. Select the Main Menu quick key → from the Configuration column select
Workflow → input the required password → select .
2. Select on the right of the workflow to be set. Make sure the workflow is
under Spot Check.
3. Select Parameters Setup → the NIBP tab.
4. Under Average, set Number of Measurements.
11.9.4.3 Setting Time between Readings
You can set the time duration between the starts of two consecutive measurement.
Follow this procedure:
1. Select the Main Menu quick key → from the Configuration column select
Workflow → input the required password → select .
2. Select on the right of the workflow to be set. Make sure the workflow is
under Spot Check.
3. Select Parameters Setup → the NIBP tab.
4. Under Average, set Time Between Readings.
11.9.4.4 Excluding the First Group of Readings in the Calculation
You can choose to exclude the first group of readings in the averaging calculation.
Follow this procedure:
1. Select the Main Menu quick key → from the Configuration column select
Workflow → input the required password → select .
2. Select on the right of the workflow to be set. Make sure the workflow is under
Spot Check.