13 - 1
13 Monitoring Respiration
13.1 Resp Safety Information
WARNING
The respiration measurement does not recognize the cause of apneas. It only
indicates an alarm if no breath is detected when a pre-adjusted time has
elapsed since the last detected breath. Therefore, it cannot be used for
diagnostic purpose.
If operating under conditions according to the EMC Standard IEC 60601-1-2
(Radiated Immunity 3V/m), field strengths above 3V/m may cause erroneous
measurements at various frequencies. Therefore, it is recommended to avoid
the use of electrically radiating equipment in close proximity to the
respiration measurement unit.
CAUTION
Only use parts and accessories specified in this manual.
Respiration monitoring is not for use on the patients who are very active, as
this will cause false alarms.
13.2 RR Display
The RR parameter area displays its source.
1. RR: Respiration rate per minute
2
3
4
5 65
2
11
Sourced from CO
2
Sourced from Mindray SpO
2