C - 6
C.1.6 EWS Default Settings
Setup AG Compensation 0%
Speed 6.25 mm/sec
CO2 Scale 50 mmHg
Waveform Type Draw
Item Default Setting
Item Default Setting
EWS Score NEWS2
Score Confirmation On
Auto Scoring / Off
Interval By Score
4 hrs (Score: 0)
2 hrs (Score: 1~4)
1 hr (Score: 5~6)
30 min (Score: 7~20)
Manual Data
Timeout
SpO2 4 hrs
Supp. O2 24 hrs
PR 1 hr
LOC 24 hrs