21 - 2
CAUTION
The equipment and accessories shall not be served or maintained while in
use with a patient.
If you discover a problem with any of the equipment, contact your service
personnel or Mindray.
Use and store the equipment within the specified temperature, humidity, and
altitude ranges.
When disposing of the packaging material, be sure to observe the applicable
waste control regulations and keep it out of children’s reach.
At the end of its service life, the equipment, as well as its accessories, must be
disposed of in compliance with the guidelines regulating the disposal of such
products. If you have any questions concerning disposal of the equipment,
please contact Mindray.
NOTE
If needed, contact the manufacture for circuit diagrams, component part
lists, descriptions, calibration instructions, or other information concerning
the repair of the equipment.
21.3 Maintenance and Testing Schedule
Follow the maintenance and testing schedule or local regulations to perform testing and
maintenance. Make sure to clean and disinfect the equipment before taking any tests
and maintenance
The following table lists the maintenance and testing schedule:
Test/Maintenance Item Recommended Frequency
Performance Tests
Visual inspection Every day, before first use.
Measurement module
performance test and
calibration
1. If you suspect that the measurement values are
incorrect.
2. Follow any repairs or replacement of relevant module.
3. Once a year for CO
2
tests.
4. Once every two years for other parameter module
performance tests.