15 - 7
◆ On: you need to confirm that whether the scoring result is saved or not after
the scoring is completed.
15.1.6.3 Setting the Manual Data Timeout
The manually input parameter data become invalid after a preset time. To set the
timeout period for the input data, follow this procedure:
1. From the EWS screen select Setup.
2. From the Manual Data Timeout area, select a desired parameter and set its
timeout period.
• If the data is expired and not updated, the monitor displays the
corresponding parameter score in outline font, and gives a timeout alarm.
15.1.6.4 Selecting the Default Scoring Tool
To set the default scoring tool for different patient catalog, follow this procedure:
1. Select the Main Menu quick key → from the System column select Maintenance
→ input the required password → select .
2. Select the CAA tab → select EWS tab.
3. From the Select Default Score area, set Default Adult Score, Default Ped Score,
and Default Neo Score.
15.1.7 Viewing History Scores
To view the history scores, follow this procedure:
1. Select the Review quick key → select Scoring Review tab.
2. Select PID and from the popup list select the ID of which the scores are to be
viewed.
3. Select Filter and choose a protocol if needed.
4. Select a record and then select Detail.
15.1.7.1 Managing the Scoring Tools
To manage the scoring tool, follow this procedure:
1. Select the Main Menu quick key → from the System column select Maintenance
→ input the required password → select .
2. Select the CAA tab → select EWS tab.
3. Select Manage Score.
◆ Delete local scoring tools: from the Local page, delete the unnecessary
scoring tools.