Mindray VS 9 - BP Averaging Display; Performing BP Averaging

Mindray VS 9
278 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11 - 12
4. Under Average, enable BP Averaging.
11.9.2 BP Averaging Display
The following figure shows the BP averaging screen.
1. Time of last measurement
2. NIBP unit
3. Select this area to set the patient position and measurement site
4. Averaged blood pressure: displayed in the form of Sys/Dia (Mean)
5. Measurements used for averaging: if the first group of measurements is displayed
with a strikethrough, it is excluded in the averaging. For details, see
11.9.4.4 Excluding the First Group of Readings in the Calculation.
6. Start NIBP measurement
11.9.3 Performing BP Averaging
Follow this procedure:
1. Prepare the patient and place the NIBP cuff as instructed in 11.6 Preparing for NIBP
Measurement.
2. Swipe on the NIBP area to the left or right to switch to BP averaging screen.
3. Select to set the patient position and measurement site.
4. Select to start the first measurement.
The monitor starts the first measurement immediately or after the set delay.
For details, see 11.9.4.1 Setting the Delay before Starting the First
Measurement.
After the first measurement, the monitor automatically takes several times
of BP measurement (see 11.9.4.2 Setting the Number of Measurement) at the
set interval (see 11.9.4.3 Setting Time between Readings).
5. At the completion of automatic measurement, select Return to check the
readings of each measurement and the averaged BP values.
2
3
1
4
5
6