C - 1
C Default Settings
C.1 Parameters Default Settings
C.1.1 SpO2 Default Settings
Item Default Setting
Alarm SpO2 On/Off On
High Adult and pediatric: 100%
Neonate: 95%
Low 90%
Priority Med
SpO2 Desat On/Off On
Low 80%
Priority High
NIBP Simul Off
Setup Sat-Seconds (for Nellcor SpO
2
)Off
Sensitivity (for Mindray SpO
2
)Med
Sensitivity (for Masimo SpO
2
)APOD
Averaging (for Masimo SpO
2
)8 s
Display PI (for Mindray SpO
2
, Masimo SpO
2
)On
Speed 25 mm/sec