11 - 16
3. Select Parameters Setup → the NIBP tab.
4. Under Orthostatic BP Measurement, set Lying Duration.
11.10.4.2 Setting Standing BP Measurement Interval
Follow this procedure:
1. Select the Main Menu quick key → from the Configuration column select
Workflow → input the required password → select .
2. Select on the right of the workflow to be set. Make sure the workflow is
under Spot Check.
3. Select Parameters Setup → the NIBP tab.
4. Under Orthostatic BP Measurement, set Standing BP Measurement Interval.
11.10.4.3 Setting the Times of Standing BP Measurement
Follow this procedure:
1. Select the Main Menu quick key → from the Configuration column select
Workflow → input the required password → select .
2. Select on the right of the workflow to be set. Make sure the workflow is
under Spot Check.
3. Select Parameters Setup → the NIBP tab.
4. Under Orthostatic BP Measurement, set Maximum Standing BP Measure-
ments.
11.10.4.4 Setting Orthostatic BP Symptoms
Follow this procedure:
1. Select the Main Menu quick key → from the System column select Maintenance
→ input the required password → select → Module.
2. Under Orthostatic BP Symptoms, select to edit a symptom. You can also
select Add Symptom to create a new symptom and then select to edit it.
NOTE
• You can set up to ten symptoms.
11.10.4.5 Setting Orthostatic BP Evaluation Criterion
You can set criteria for evaluation of the pressure differences. A negative/positive result
displays on the screen at the completion of the measurement.
Follow this procedure: