A - 13
Measurement ranges
(mmHg)
Adult Pediatric Neonate
Systolic: 25 to 290 25 to 240 25 to 140
Diastolic: 10 to 250 10 to 200 10 to 115
Mean: 15 to 260 15 to 215 15 to 125
Accuracy Max mean error: ±5 mmHg
Max standard deviation: 8 mmHg
Static pressure
measurement range
0 mmHg to 300 mmHg
Static pressure
measurement accuracy
±3 mmHg
Resolution 1 mmHg
Initial cuff inflation
pressure range (mmHg)
(Deflation algorithm)
Adult:
Pediatric:
Neonate:
80 to 280
80 to 210
60 to 140
Default initial cuff
inflation pressure
(mmHg) (Deflation
algorithm)
Adult:
Pediatric:
Neonate:
160
140
90
Software overpressure
protection
Adult:
Pediatric:
Neonate:
297±3 mmHg
297±3 mmHg
147±3 mmHg
Hardware overpressure
protection
Adult:
Pediatric:
Neonate:
≤330 mmHg
≤330 mmHg
≤165 mmHg
Alarm limit Range (mmHg) Step (mmHg)