3 - 15
SpO2 Sensitivity For Masimo SpO
2
 modules, see 8.6.4 Setting 
SpO
2
 Sensitivity (for Masimo SpO
2
). For 
Mindray SpO
2
 modules, see 8.6.8 Changing 
the Sensitivity (for Mindray SpO
2
).
Both
Display PI Sets whether to display PI in the SpO
2
 
parameter area.
Speed Sets the sweep speed of Pleth waveforms.
NIBP Simul Switch on for patients performing SpO
2
 and 
NIBP monitoring on the same limb. The SpO
2
 
alarm status is locked until the NIBP 
measurement ends.
Continuous 
Monitoring
PR Pulse Volume Sets the pulse volume. Both
Irr. PR Switch Selects whether to enable the Irregular PR 
function. If enabled, you can mark the pulse 
as regular or irregular on the PR screen.
Spot Check
Temp Measurement Site 
(Manual)
Sets the default measurement site for Temp 
measurement.
Both
RR Timer Reminder 
Interval
Sets the interval of reminder tone for the 
timer used for manual RR. For details, see 
13.5.5 Setting Timer Interval.
Both
Apnea Delay 
(Adult)/(Ped)/
(Neo)
Sets the delay for giving an Apnea alarm. Continuous 
Monitoring
Tab Menu Item Description Work Mode