3 - 21
■ The alarm setup menus will not be displayed.
■ Alarm limits cannot be set.
Different features between continuous monitoring mode and spot check mode:
3.10.1.3 Standby Mode
You can temperately stop patient monitoring without switching off the monitor by
entering the standby mode. In standby mode, the monitor behaves as follows:
■ Stops all parameter measurement.
■ Disables all the alarms and prompt messages, except for the battery low alarm.
■ Turns screen brightness to the dimmest after entering the standby mode.
You can switch the monitor to standby mode manually, or set Auto Enter Standby for
the monitor to enter standby mode automatically.
To switch the monitor to standby mode:
1. Press the power switch, or select the Main Menu quick key → from the Patient
Management column select Standby.
2. If the monitor is in Continuous Monitoring mode, select a location in the drop
down list to define where the patient is when the monitor enters the standby
mode. If the monitor is in Spot Check mode, skip this step.
3. Select OK.
To set Auto Enter Standby:
Functions Continuous Monitoring mode Spot check mode
Configure and use Sat-Seconds
(Nellcor)
Yes No
Access Alarm Setup tab Yes No
Access Events tab Yes No
Access Graphic Trends tab Yes No
Auto data updating with the
central monitoring system
Yes No
BP Averaging No Yes
Orthostatic BP measurement No Yes
Measure CO
2
(VS 9 series) Yes No
SpO
2
sourced RR (Mindray SpO
2
module)
No Yes