DMTA-10084-01EN, Rev. 2, November 2018
Calibration
95
Figure 8-3 Calibration Dialog box initial setup
7. Adjust G1Start until the gate is over the first back wall echo.
8. Press the Tab key to highlight the Gain adjustment box, and then complete the
Auto 80% procedure (see Figure 8-4 on page 96). See “Using the Auto XX%
Feature” on page 60 for details on the Auto XX% procedure.